You are here: Installation and Maintenance Guide > Calibration > Calibration Procedures > Linearity Correction

Linearity Correction

Most users will scan with the system set up for Closed Loop feedback. However, if you are running applications in Open Loop that demand good linearity, the following procedure can be used to optimize the linearity correction parameters for individual scanners. As discussed previously, linearity correction is especially important for long-range scanners.

Adjust Mag0 and Arg

Set the mag0 and arg values while noting the scan beginning and ending.

NOTE: When Line direction is set to Trace, the beginning of the fast scan is on the left, as indicated by the arrow base. Because the monitor screen itself may be nonlinear, use the Realtime Zoom box to check mag0 and arg.
NOTE: When viewing up or down scans, check the image against itself, not the residual image from the previous scan. Be careful to compare only the part of the scan drawn since the last parameter change.
Adjust Fast Mag0
Adjust Fast Arg
Adjust Slow Mag0
Adjust Slow Arg

Adjust Fast Mag1

Initial Adjustment
Fine Adjustment

Adjust Slow Mag1

 
  1. With the Scan size set to 150 V, select Tools > Calibrate > X-Y...
 
  1. Select Slow mag1 and input the value from Fast mag1.
NOTE: For medium-sized scanners (C to F), further adjustments should not be necessary. The Slow mag1 value is usually 100–120 percent of the Fast mag1 value.
 
  1. If greater precision is desired, adjust Slow mag1 using the same procedure for adjusting Slow mag0.
 
  1. Set the Zoom box for the beginning of the scan and then check against the ending. Because the scan is small, use a Zoom box up to one-half as large as the scan.
 
  1. If the end of the scan is larger than the beginning, decrease the Slow mag1 value. If the end is too small, increase the value of Slow mag1.
 
  1. Wait one complete frame with the new value before readjusting the Slow mag1 value.
 
  1. Check the final result by capturing an image and checking it with the Offline > Modify > Zoom window. The scanner is now ready for calibration of the X and Y parameters.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.